                  ▷Specification | 
                   
                  ◼ HCHO
  detection technique: Fluorimetric, using Hantzsch (acetyl-acetone) reaction 
                  ◼ Linear
  detection range: 0.1ppb to 3000ppb (gaseous), 150ng/liter - 5mg/liter
  (liquid) 
                  ◼ Detection
  limit: <100ppt (gaseous), <150ng/liter eq. <2x10^ 
                  (-9) molar (liquid) 
                  ◼ Sample gas
  temperature: >0°C to +120°C 
                  ◼ Calibration
  and zeroing: Automatic zeroing and semi-automatic calibration using liquid
  standards or automatic using internal gas generator (optional) 
                  ◼ Weight and
  dimension: 20kg, fit for 19" rack (whd: 45cm×15cm×56cm) 
                  ◼ Power
  requirements: 110VAC / 220VAC, 110W, 24VDC on request 
                   | 
| 
                  ▷Features | 
                    
                  ◼ Continuous
  online monitoring of formaldehyde emissions with unique sensitivity of 100ppt 
                  ◼ Analysis of
  gaseous and liquid samples with only one instrument 
                  ◼ Designed for
  climate research, environmental air monitoring and indoor air quality control 
                  ◼ Ideal for
  emission control of wood-, plastic-, and fabric based products 
                  ◼ Can be
  employed together with up to 16 emission test chambers 
                   | 
| 
                  ▷Applications | 
                   
                  ◼ Climate and environmental research / Trace
  gas analysis 
                  ◼ Online monitoring of HCHO emission /
  Emission limit control 
                  ◼ Process control / Quality control for the
  chemical industry 
                  ◼ Chamber testing of wood-, plastic-,
  fabric-based products using up to 12 chambers simultaneously 
                  ◼ Measurement of indoor air quality 
                  ◼ Replacement for the wet chemistry in the
  acetyl-acetone method 
                   |